Related Experiment Video
Updated: Nov 30, 2025

Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry UPLC-MS
Published on: March 14, 2013
Development of a ultra-performance LC-MS/MS method for quantification of GW788388 and its applications in
Lina Fang1, Chunling Zhou2, Meihua Jiang1
1College of Pharmacy, Shenyang Medical College, Shenyang 110034, PR China.
Abstract:
Aim: GW788388 is a selective and orally active TGF-β1 receptor inhibitor that shows potent activity in renal fibrosis. We aimed to establish and validate a simple and sensitive ultra-performance LC-MS/MS method for the determination of GW788388 in plasma samples. Methodology & results: GW788388 in rat plasma was processed with protein precipitation method and then separated on a C18 column. The calibration curve presented a good linearity in the range of 1.0-1200 ng/ml, with satisfactory accuracy (relative error, [-17.5% < relative error <11.7%) and precision (CV <8.9%) for all quality control samples. After oral administration, GW788388 was absorbed quickly and reached a peak concentration of 595.3 ± 60.2 ng/ml after 20 min. Conclusion: The validated method provides a quantification method of GW788388 in rat plasma in detail, and can be utilized to successfully describe the pharmacokinetic profile of GW788388.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
10:13Author Spotlight: An Efficient Methodology to Confidently Differentiate and Characterize Fentanyl Analogs
Published on: November 8, 2024