Breath emulator for simulation and modelling of expired tidal breath carbon dioxide characteristics
Lewis Fleming1, Des Gibson1, David Hutson1
1Institute of Thin Films, Sensors and Imaging, School of Engineering and Computing, University of the West of Scotland, PA1 2BE Paisley, Scotland, UK.
Background:
In this work we describe a breath emulator system, used to simulate temporal characteristics of exhaled carbon dioxide (CO2) concentration waveform versus time simulating how much CO2 is present at each phase of the human lung respiratory process. The system provides a method for testing capnometers incorporating fast response non-dispersive infrared (NDIR) CO2 gas sensing devices - in a clinical setting, capnography devices assess ventilation which is the CO2 movement in and out of the lungs. A mathematical model describing the waveform of the expired CO2 characteristic and influence of CO2 gas sensor noise factors and speed of response is presented and compared with measured and emulated data.
Objective:
A range of emulated capnogram temporal waveforms indicative of normal and restricted respiratory function demonstrated. The system can provide controlled introduction of water vapour and/ or other gases, simulating the influence of water vapour in exhaled breath and presence of other gases in a clinical setting such as anaesthetic agents (eg N2O). This enables influence of water vapour and/ or other gases to be assessed and modelled in the performance of CO2 gas sensors incorporated into capnography systems. As such the breath emulator provides a means of controlled testing of capnometer CO2 gas sensors in a non-clinical setting, allowing device optimisation before use in a medical environment.
Methods:
The breath emulator uses a unique combination of mass flow controllers, needle valves and a fast acting switchable pneumatic solenoid valve (FASV), used to controllably emulate exhaled CO2 temporal waveforms for normal and restricted respiratory function. Output data from the described emulator is compared with a mathematical model using a range of input parameters such as time constants associated with inhalation/ exhalation for different parts of the respiratory cycle and CO2 concentration levels. Sensor noise performance is modelled, taking into account input parameters such as sampling period, sensor temperature, sensing light throughput and pathlength.
Results:
The system described here produces realistic human capnographic waveforms and has the capability to emulate various waveforms associated with chronic respiratory diseases and early stage detection of exacerbations. The system has the capability of diagnosing medical conditions through analysis of CO2 waveforms. Demonstrated in this work the emulator has been used to test NDIR gas sensor technology deployed in capnometer devices prior to formal clinical trialling.
More Related Videos
08:34Investigation into Deep Breathing through Measurement of Ventilatory Parameters and Observation of Breathing Patterns
Published on: September 16, 2019
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Mechanism of Breathing II: Expiration
Expiration, or exhaling, is a complex physiological process that begins as the inspiratory muscles begin to relax. This relaxation triggers a series of events that epitomize the efficiency of the respiratory system.
Mechanism of Expiration:
Respiratory Volumes and Capacities
Respiratory Volumes and Capacities I
Breathing
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.
The Functional Residual Capacity (FRC) represents the air in the...