Evaluation of Temperature Distribution Around the Probe in Cryoablation of Lipiodol-Mixed-Tissue Phantom
Masanori Yamashita1, Hiroshi Miura2, Yu Ohara2
1Department of Radiology, Kyoto Prefectural University of Medicine, 465, Kajii-cho, Kamigyo-ku, Kyoto, 602-8566, Japan. m-yama@koto.kpu-m.ac.jp.
Purpose:
To determine whether lipiodol, which has low thermal conductivity, influences ice ball formation during cryoablation of a lipiodol-mixed-tissue phantom.
Materials And Methods:
Lipiodol-mixed-tissue phantoms were created by injecting lipiodol (4-6 ml) into the renal arteries of ex vivo porcine kidneys (lipiodol group). A cryoprobe (CryoHit™ Needle S) with a holder that was set with thermocouples at various positions around the cryoprobe was inserted. After freezing for 300 s, the followings were evaluated: ice ball size on CT, temperature distribution around the cryoprobe, and calculated distances at 0 °C and - 20 °C. Each variable was compared between lipiodol group (n = 6) those obtained in a control group without lipiodol injection (n = 6).
Results:
Mean ice ball diameter (width/length) on CT was 22.1 ± 2.3/22.9 ± 2.3 mm in the lipiodol group and 21.6 ± 0.7/22.2 ± 1.3 mm in the control group. Mean cryoprobe temperature was - 118 ± 3.0 °C in the lipiodol group and - 117 ± 2.6 °C in the control group. In both groups, temperature at the 3 mm thermocouple reached approximately - 50 °C and was < 0 °C within ~ 10 mm of the cryoprobe. Temperature of 0/- 20 °C occurred at a mean distance from the cryoprobe of 11.1 ± 0.5/6.9 ± 0.4 mm in the lipiodol group and 11.0 ± 0.2/6.9 ± 0.2 mm in the control group. There was no significant difference in any variable between the groups.
Conclusion:
The inclusion of lipiodol in a tissue phantom had no negative effects on ice ball formation that were related to thermal conductivity.
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...