Related Experiment Video
Updated: Nov 29, 2025

Subdural Soft Electrocorticography ECoG Array Implantation and Long-Term Cortical Recording in Minipigs
Published on: March 31, 2023
Development of Neuromonitoring Pedicle Screw - Results of Electrical Resistance and Neurophysiologic Test in Pig
Woo-Seok Bang1, Jonghoo Park2, Kyoung-Tae Kim3
1Department of Neurosurgery, Topspine Hospital, Daegu, Korea.
Objective:
To analyze the electrical resistance of a newly developed neuromonitoring pedicle screw (Neuro-PS) and to verify the electrophysiologic properties of the Neuro-PS in a pig model.
Methods:
We developed 2 types of the Neuro-PS in which a gold lead was located internally (type I) and externally (type II). We measured the electrical resistance of the Neuro-PS and the conventional screw and analyzed the electrical thresholds of triggered EMG (t-EMG) of each screw by intentionally penetrating the medial pedicle wall and contacting the exiting nerve root in a pig model.
Results:
The electrical resistances of the Neuro-PS were remarkably lower than that of the conventional screw. In electrophysiologic testing, only the type II Neuro-PS under the leadnerve contact condition showed a significantly lower stimulation threshold as compared to the conventional screw.
Conclusion:
The Neuro-PS demonstrated lower electrical resistances than the conventional screw. The type II Neuro-PS under the lead-nerve contact condition showed a significantly lower stimulation threshold compared to that of the other screws in the t-EMG test.