Integrating photoacoustic tomography into a multimodal automated breast ultrasound scanner
Corey Kelly1, Amir Refaee1, Septimiu E Salcudean1
1University of British Columbia, Department of Electrical and Computer Engineering, Vancouver, Britis, Canada.
Significance:
Photoacoustic tomography (PAT) is a promising emergent modality for the screening and staging of breast cancer. To minimize barriers to clinical translation, it is common to develop PAT systems based upon existing ultrasound hardware, which can entail significant design challenges in terms of light delivery. This often results in inherently non-uniform fluence within the tissue and should be accounted for during image reconstruction.
Aim:
We aim to integrate PAT into an automated breast ultrasound scanner with minimal change to the existing system.
Approach:
We designed and implemented an illuminator that directs spatially non-uniform light to the tissue near the acquisition plane of the imaging array. We developed a graphics processing unit-accelerated reconstruction method, which accounts for this illumination geometry by modeling the structure of the light in the sample. We quantified the performance of this system using a custom, modular photoacoustic phantom and graphite rods embedded in chicken breast tissue.
Results:
Our illuminator provides a fluence of 2.5 mJ cm - 2 at the tissue surface, which was sufficient to attain a signal-to-noise ratio (SNR) of 8 dB at 2 cm in chicken breast tissue and image 0.25-mm features at depths of up to 3 cm in a medium with moderate optical scattering. Our reconstruction scheme is 200 × faster than a CPU implementation; it provides a 25% increase in SNR at 2 cm in chicken breast tissue and lowers image error by an average of 31% at imaging depths >1.5 cm compared with a method that does not account for the inhomogeneity of the illumination or the transducer directivity.
Conclusions:
A fan-shaped illumination geometry is feasible for PAT; however, it is important to account for non-uniform fluence in illumination scenarios such as this. Future work will focus on increasing fluence and further optimizing the ultrasound hardware to improve SNR and overall image quality.
Related Concept Videos
Imaging Studies II: Ultrasonography
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...