Related Experiment Video
Updated: Nov 29, 2025

An Open-Source Normothermic Perfusion System Designed for Research Scientists
Published on: July 18, 2025
An easy disinfection strategy for pipes connecting hemodialysis equipment
Masashi Tagaya1, Yosuke Oda1, Aki Kimura1
1Department of Medical Engineering, National Hospital Organization Kure Medical Center, Kure, Hiroshima, Japan.
Introduction:
A hemodialysis room has pipes connecting the console and central fluid equipment. While these pipes require disinfection, reports detailing disinfection strategies are unavailable. Therefore, we aimed to compare two easy disinfection strategies differing in sanitization frequency and sanitizer concentration.
Methods:
Reverse osmosis water (ROW) purifying equipment and six dialysis consoles were connected by 20 m of pipes. Only ROW flowed through these pipes, because the dialysate solution was constituted at each console. The pipes were sanitized by two strategies: (1) strong and monthly (hypochlorite concentration: 100 ppm) or (2) weak and weekly (5 ppm). Both strategies were easy because the sodium hypochlorite was simply added to the ROW tank. To estimate sanitization efficacy, endotoxin counts at the ROW tank outlet, the end of the pipe, and the pipe after the endotoxin-cutting filter in each console were measured monthly for six continuous months. These counts were compared between the two sanitization strategies.
Results:
The endotoxin counts at the ROW tank outlet and the end of the pipe were 0.004-0.017 and 0.012-0.081 EU/mL, respectively, in the strong and monthly strategy, and 0.001-0.003 and 0.001-0.005 EU/mL, respectively, in the weak and weekly strategy. The endotoxin counts at the pipe after the endotoxin-cutting filter were less than 0.001 EU/mL during the study period in both strategies.
Conclusion:
A weekly disinfection strategy was more effective than a monthly one, despite the lower hypochlorite concentration. The present study suggests that frequency is the most important factor in the disinfection of pipes in a dialysis room.
More Related Videos
12:11A Modified EPA Method 1623 that Uses Tangential Flow Hollow-fiber Ultrafiltration and Heat Dissociation Steps to Detect Waterborne Cryptosporidium and Giardia spp.
Published on: July 9, 2012
17:30Pseudomonas aeruginosa and Saccharomyces cerevisiae Biofilm in Flow Cells
Published on: January 15, 2011
Related Concept Videos
Hemodialysis I: Introduction
Hemodialysis II: Procedure and Complications
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Extracorporeal Removal of Drugs: Peritoneal Dialysis and Hemodialysis