Smaranda Buduru1, Silvia Balhuc1, Alexandru Ciumasu1
1Department of Prosthodontics, Faculty of Dental Medicine, Iuliu Hatieganu University of Medicine and Pharmacy, Cluj-Napoca, Romania.
Background And Aims:
The objective of this study was to evaluate the accuracy of the CADIAX Compact 2 axiograph in confirming the presumptive clinical diagnosis of temporomandibular dysfunction, established according to the muscular-articular clinical examination.
Methods:
A retrospective analytical study was carried out on 50 patients, aged between 21 and 62 years, 31 women and 19 men. A clinical diagnosis was established after masticatory muscles, temporomandibular joints and mandibular border limit movements were assessed. During the same day a computerized axiography (Cadiax Compact 2, Gamma, Vienna, Austria) was performed in order to confirm the diagnosis.
Results:
The study had a sensitivity of 100%, which means that the CADIAX device detected all clinically positive patients. On the other hand, the specificity of the study was 74%, representing the percentage of patients negatively diagnosed by CADIAX device out of the total number of patients clinically diagnosed as negative. The 88% accuracy shows the proportion in which the CADIAX device provides a diagnosis equal to the clinical one.
Conclusions:
Of all the methods for investigating temporomandibular joint dysfunction, axiography is less invasive, does not irradiate the patient, it is relatively easy to use and it offers functional information about the TMJ function.