Related Experiment Video
Updated: Nov 29, 2025

A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Importance of the epiphyseal ring in OLIF stand-alone surgery: a biomechanical study on cadaveric spines
Xuyang Zhang1,2, Hao Wu3, Yilei Chen1,2
1Department of Orthopaedics, Sir Run Run Shaw Hospital, Zhejiang University School of Medicine, No. 3, Qingchun Road East, Hangzhou, 310016, People's Republic of China.
Purposes:
To explore the function of endplate epiphyseal ring in OLIF stand-alone surgery using a biomechanical model to reduce the complications of endplate collapse and cage subsidence.
Methods:
In total, 24 human cadaveric lumbar function units (L1-2 and L3-4 segments) were randomly assigned to two groups. The first group was implanted with long fusion cages which engaged with both inner and outer regions of epiphyseal ring (Complete Span-Epiphyseal Ring, CSER). Those engaged with only the inner half of epiphyseal ring were the second group (Half Span-Epiphyseal Ring, HSER). Each group was divided into two subgroups [higher cage-height (HH) and normal cage-height (NH)]. Specimens were fixed in testing cups and compressed at approximately 2.5 mm/s, until the first sign of structural failure. Trabecular structural damage was analyzed by Micro-CT, as well as the difference of bone volume fraction (BV/TV), trabecular thickness (Tb.Th) et al. in different regions.
Results:
Endplate collapse was mainly evident in the inner region of epiphyseal ring, where trabecular injury of sub-endplate bone was most concentrated. Endplate collapse incidence was significantly higher in HSER than CSER specimens (P = 0.017). A structural failure occurred at a lower force in HSER (1.41 ± 0.34 KN) compared with CSER (2.44 ± 0.59 KN). HH subgroups failed at a lower average force than NH subgroups. Micro-CT results showed a more extensive trabecular fracture in HSER specimens compared to CSER specimens, especially in HH subgroup.
Conclusions:
Endplate collapse is more likely to occur with short half span cages than complete span cages, and taller cages compared with normal height cages. During OLIF surgery, we should choose cages matching intervertebral disc space height and place the cages spanning over the whole epiphyseal ring to improve support strength.
More Related Videos
Related Concept Videos
Changes in the Appendicular Skeleton with Age
Initially, the limb buds consist of a core of mesenchyme covered by a layer of ectoderm. The ectoderm at the end of the limb bud thickens to form a narrow crest called the apical ectodermal ridge. This ridge stimulates the underlying...
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...