Evaluation of PASCO MIC-ID system for identifying gram-negative bacilli
D L Rhoden1, B Schable, P B Smith
1Nosocomial Infections Laboratory Branch, Centers for Disease Control, Atlanta, Georgia 30333.
Journal of Clinical Microbiology
|December 1, 1987
Summary
The semi-automated PASCO MIC-ID system accurately identified 90.9% of gram-negative bacilli cultures. Improvements to its database could enhance the identification of less common bacterial species.
Area of Science:
- Clinical Microbiology
- Bacteriology
- Diagnostic Systems Evaluation
Background:
- Accurate identification of gram-negative bacilli is crucial for effective patient treatment and infection control.
- Semi-automated systems offer potential for improved efficiency in clinical microbiology laboratories.
Purpose of the Study:
- To evaluate the accuracy and performance of the production model of the semi-automated PASCO MIC-ID system.
- To assess the system's identification capabilities across various groups of gram-negative bacilli, including fermenters and nonfermenters.
Main Methods:
- Evaluation of 536 cultures from 122 groups/species of gram-negative bacilli (typical and atypical strains).
- Utilized serologic and additional biochemical tests as recommended by the PASCO system to supplement identification.
- Analysis of identification accuracy rates for nonfermenters, enteric fermenters, and oxidase-positive fermenters.
Main Results:
- The PASCO system achieved an overall accurate identification rate of 90.9% for 536 cultures.
- Specific accuracy rates included 90.8% for nonfermenters, 93.8% for enteric fermenters, and 78.9% for oxidase-positive fermenters.
- The system identified commonly encountered organisms with ≥95% accuracy; common misidentifications were noted for specific species.
Conclusions:
- The PASCO MIC-ID system demonstrates good overall accuracy for gram-negative bacilli identification, particularly for common organisms.
- Minor improvements to the system's database could address identification challenges with less frequent or atypical strains.
- The system is user-friendly, with straightforward inoculation and reading procedures.