Related Experiment Video
Updated: Nov 28, 2025

A Computer-assisted Multi-electrode Patch-clamp System
Published on: October 18, 2013
Design and experimental evaluation of an automated catheter operating system
Somayeh Norouzi-Ghazbi1, Ali Mehrkish2, Ibrahim Abdulhafiz1
1Biomedical Engineering, Ryerson University, Toronto, ON, Canada.
This study introduces Althea II, a novel robot-assisted catheterization system (RACS) designed to overcome limitations of current technologies. Althea II enhances accuracy and precision in catheter-based interventions while reducing X-ray exposure for medical professionals.
Area of Science:
- Medical Robotics
- Interventional Cardiology
- Surgical Navigation
Background:
- Manual catheter-based interventions (CBIs) pose risks to interventionalists due to X-ray exposure and are limited by operator expertise and fatigue.
- Existing robot-assisted catheterization systems (RACS) have drawbacks including instrument specificity, lack of portability, and limited operational modes.
Purpose of the Study:
- To develop an improved RACS, named Althea II, addressing the shortcomings of previous systems and enhancing efficiency and safety in CBIs.
- To create a versatile and user-friendly system for both research and clinical applications, including cardiovascular interventions.
Main Methods:
- Development of Althea II with a novel, lightweight structural design for broader instrument compatibility.
- Integration of a tip detection algorithm into the graphical user interface (GUI) for image-based and fully automatic navigation.
- Evaluation of system performance, including accuracy, precision, assembly time, and user-friendliness.
Main Results:
- Althea II achieved over 90% accuracy in navigating the catheter tip to a target in both velocity and positioning modes.
- The integrated tip detection algorithm in the upgraded GUI demonstrated an accuracy of 0.05 mm.
- Assembly time was reduced fivefold to 20 minutes, and the system offers multiple operation modes, portability, and remote functionality.
Conclusions:
- Althea II represents a significant advancement in RACS, offering improved accuracy, precision, and reduced X-ray exposure in catheter-based interventions.
- The system's design, including its independency from specific catheters and enhanced GUI, makes it suitable for beginners and various clinical applications.
- Preliminary studies confirm Althea II's accuracy, repeatability, feasibility, and superior setup capabilities compared to existing RACS.
More Related Videos
05:57A Teleoperated Robotic System-Assisted Percutaneous Transiliac-Transsacral Screw Fixation Technique
Published on: January 6, 2023
09:05Catheterization of the Carotid Artery and Jugular Vein to Perform Hemodynamic Measures, Infusions and Blood Sampling in a Conscious Rat Model
Published on: January 30, 2015