False Asystole Alarms Post-Temporary Pacemaker Placement Due to Pseudo-fusion
Alyssa M Vermeulen1, Frank Zimmerman1, Hoang H Nguyen2,3
1Department of Pediatrics, Rush University Medical Center, Chicago, IL, 60612, USA.
Abstract:
An infant with congenital heart block and hemodynamically significant bradycardia underwent therapeutic temporary pacing wires placement. Post-operatively, frequent "asystole" alarms were observed on telemetry causing distress to both the family and the nursing staff. Investigation of these alarms showed that pacemaker malfunction led to monitor pseudo-malfunction. The alarms were alleviated with mindful setting of the pacemaker and telemetry monitor parameters. This case highlights the challenges of pacemaker placement and monitoring of very small infants in the intensive care setting. Awareness of these challenges would help in troubleshooting pacemaker and telemetry monitor issues.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Cardiopulmonary Resuscitation III: AED Use
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Mechanism of Cardiac Arrhythmias