Evaluation of an accelerated 3D modulated flip-angle technique in refocused imaging with an extended echo-train
1Department of Radiology, Yueyang Hospital of Integrated Traditional Chinese and Western Medicine, Shanghai University of Traditional Chinese Medicine, Shanghai, China.
Aim:
To accelerate the acquisition of high-resolution magnetic resonance imaging (MRI) by using the three-dimensional (3D) matrix sequence with compressed sensing and to compare it with conventional two-dimensional (2D) proton-density (PD) and fast spin-echo (FSE) sequences.
Materials And Methods:
3D matrix, 2D FSE, and PD sequences were acquired from 68 participants using 3 T magnetic resonance imaging (MRI). Two radiologists scored image quality independently on a four-point scale. The structural similarity index (SSIM), and signal- (SNRs) and contrast-to-noise ratios (CNRs) of different anatomical structures of the knee were assessed and compared between sequences using Wilcoxon signed-rank tests and Cohen's kappa.
Results:
The median acquisition time reduction was 44.5%. There was a substantial to perfect agreement for the rating between the 3D matrix FSE and 2D FSE or PD sequences when evaluating cartilage, subchondral bone, and ligaments (κ=0.783-872, p>0.05). The mean SSIM values between the 3D matrix FSE and 2D FSE, and between the 3D matrix PD and 2D PD sequences was 0.994 and 0.971, respectively, which are acceptable. No significant differences were found in SNR between the 3D matrix FSE and 2D FSE, and between the 3D matrix PD and 2D PD sequences, even though the SNR appeared to be higher on routine 2D sequences. The CNR of subchondral bone-meniscus, subchondral bone-joint fluid, and meniscus-joint fluid did not differentiate significantly between the 3D matrix sequence and routine 2D sequences.
Conclusions:
3D matrix reduced the acquisition time in routine clinical knee MRI without the loss in image quality, SNR, and CNR.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...