Monitoring oxygen saturation and heart rate during neonatal transition. comparison between two different pulse
Rasha Khoury1, Gil Klinger2,3, Yakir Shir4
1Department of Pediatrics, Schneider Children's Medical Center of Israel, Petach Tikvah, Israel.
Objective:
To compare the efficacy and reliability of two pulse-oximeters (POx) (Masimo Radical-7 and Nellcor™ Oxymax Bedside) and evaluate the feasibility of routine ECG monitoring during delivery room transition.
Study Design:
Prospective observational comparative study. Sixty newborns were connected simultaneously to both POxs and ECG monitor (as a gold standard for HR). Times to achieve a stable signal were compared. Heart rates were compared to simultaneous ECG.
Results:
A significant difference in times to stable signal was found: Mean, Median (Interquartile range) for Nellcor and Masimo, were 15, 8.5 (6-18) and 27, 12 (9-34) seconds, respectively. Compared to ECG, false bradycardia was displayed in 18 of 55 (35%) newborns by the Masimo POx and in no newborns by the Nellcor POx. Attaching the ECG monitor was feasible but consumed additional resources.
Conclusions:
The time for achievement of a stable saturation reading in an uncomplicated resuscitation setting differed significantly between POxs.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Special considerations while measuring pulse