Two novel protocols for cleaning residual simethicone and fluid in patient-ready duodenoscopes
Chen-Ling Peng1, Tao-Chieh Liu2, Hsiu-Po Wang3
1Department of Integrated Diagnostics and Therapeutics, National Taiwan University Hospital, National Taiwan University College of Medicine, Taipei, Taiwan.
Background And Aim:
Approximately 42-95% of working channels have been reported to show the presence of residual fluid despite endoscope reprocessing. The aim of this study was to design two novel protocols for cleaning residual simethicone and demonstrate its efficiency by evaluating the residual fluid and cleanliness in the working channels of patient-ready duodenoscopes.
Methods:
The designed protocol for cleaning residual simethicone was implemented in manual cleaning and/or high-level disinfection (HLD). The residual fluid inside the working channels was estimated by visual inspection. Adenosine triphosphate (ATP) values were evaluated to determine cleanliness after manual cleaning.
Results:
Manual cleaning with novel simethicone cleaning protocol demonstrated a significant decrease in fluid droplets (14.6 ± 29.9 vs 0 ± 0, P < 0.001) and ATP values (157 ± 196 relative light units [RLUs] vs 52 ± 41 RLUs, P = 0.031). HLD with simethicone cleaning protocol, using either enzymatic detergent with effective for cleaning simethicone or cleaning time set in the automatic endoscope reprocessor program for 8 min, demonstrated significant decrease in the number of fluid droplets. Follow-up after the implementation of the simethicone cleaning protocol showed a significant decrease in fluid droplets (37.4 ± 41.0 vs 2.1 ± 5.5, P = 0.003) and ATP values (271 ± 268 RLUs vs 82 ± 136 RLUs, P = 0.021).
Conclusions:
Simethicone cleaning protocol is advantageous for significantly decreasing fluid droplets and ATP values within endoscope working channels. After manual cleaning with the simethicone cleaning protocol, in particular, no retained fluid droplet was observed in patient-ready duodenoscopes.
More Related Videos
10:29Semi-automatic PD-L1 Characterization and Enumeration of Circulating Tumor Cells from Non-small Cell Lung Cancer Patients by Immunofluorescence
Published on: August 14, 2019
09:30Author Spotlight: Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
Related Concept Videos
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Hand hygiene
Hand washing...
Endoscopic Procedures V: ERCP
Patient...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...