Reduction of scatter in breast CT yields improved microcalcification visibility
1Malcova LLC, 3000 Falls Rd Suite 400, Baltimore, MD 21211, United States of America.
Abstract:
The inadequate visibility of microcalcifications-small calcium deposits that cue radiologists to early stages of cancer-is a major limitation in current designs of dedicated breast computed tomography (bCT). This limitation has previously been attributed to the constituent components, spatial resolution, and utilized dose. Scattered radiation has been considered an occurrence with low-frequency impacts that can be compensated for in post-processing. We hypothesized, however, that the acquisition of scattered radiation has a far more detrimental impact on clinically relevant features than has previously been understood. Critically, acquisition of scatter leads to the reduced visibility of microcalcifications. This hypothesis was investigated and supported via mathematical derivations and simulation studies. We conducted a series of comparative studies in which four bCT systems were simulated under iso-dose and iso-resolution conditions, characterizing the dependencies of microcalcification contrast on accumulated scatter. Included among the simulated systems is a novel bCT design-narrow beam bCT (NB-bCT)-that captures nearly zero scatter. We find that current bCT systems suffer from significant levels of scatter. As validated in theory, depending on the system and size of microcalcifications, between 25% and over 70% of contrast resolution is lost due to scatter. The results in NB-bCT, however, provide evidence that by removing scatter build-up in projections, the contrast of microcalcifications in a bCT image is preserved, regardless of their size or location in the breast.
More Related Videos
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
15:48Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...