Related Experiment Video
Updated: Nov 26, 2025

A Modified Surgical Technique for Kidney Transplantation in Mice
Published on: July 22, 2022
Magnetic Ureteral Stents Are Feasible in Kidney Transplant Recipients: A Single-Center Experience
P F Pohlmann1, M Kunzelmann1, K Wilhelm1
1Department of Urology, Faculty of Medicine, University of Freiburg-Medical Centre, Freiburg, Germany.
Background:
Insertion of ureteral catheters is a common procedure in kidney transplantation. The stent is usually removed by cystoscope. Magnetic ureteral stents may be an alternative to conventional stents.
Objective:
To assess the functional efficacy and feasibility of magnetic double J (DJ) stents in kidney transplant recipients.
Methods:
We used 6 Fr (diameter), 22 cm (length) magnetic DJs. We examined 7 cases of exclusively AB0-identical living donations. Stent were removed 10-12 days after transplantation. Ureteral Stent Symptoms Questionnaire (USSQ) and visual analog scale (VAS) were used to determine quality of life and pain of the recipients. The total removal time was recorded and cost reduction was calculated.
Results:
Removal of the magnetic DJ was successful in all cases. The mean±SD duration of the removal was 3.4±1.6 min. The mean±SD overall pain score on the VAS during the procedure was 2.6±1.1. Using this technique was associated with a cost reduction of € 130.
Conclusion:
Using magnetic ureteral stents is a feasible option for living donation AB0-identical kidney transplant recipients.
Related Concept Videos
Kidney Transplant II: Surgical Procedure
Kidney Transplant I: Introduction
Kidney Transplant III: Nursing Management
Imaging Studies IV: Magnetic Resonance Imaging