Related Experiment Video
Updated: Nov 25, 2025

A Vibrotactile Feedback Device for Seated Balance Assessment and Training
Published on: January 20, 2019
Digital bathroom scales with open source software provide valid dynamic ground reaction force data for assessment and
Ross A Clark1, Benjamin F Mentiplay2, Hong Han Tan3
1School of Health and Sport Sciences, University of the Sunshine Coast, QLD, Australia.
Modified digital bathroom scales offer an affordable way to accurately measure vertical ground reaction force. These low-cost devices demonstrate excellent reliability and validity for clinical and physical capacity assessments.
Area of Science:
- Biomechanics
- Biomedical Engineering
- Rehabilitation Technology
Background:
- Accurate measurement of vertical ground reaction force (vGRF) is crucial for assessing clinical factors like weight-bearing asymmetry and rate of force development.
- Existing vGRF measurement technologies are often prohibitively expensive, limiting accessibility in various settings.
Purpose of the Study:
- To evaluate the validity and reliability of using modified, low-cost digital bathroom scales for obtaining static and dynamic vGRF data.
- To determine if inexpensive scales can serve as a viable alternative to costly force measurement systems.
Main Methods:
- Four modified digital bathroom scales (two brands) underwent mechanical loading trials (0.01-65 kg) to assess accuracy, hysteresis, and non-linearity.
- Thirty-two adults performed dynamic tasks (sit-to-stand, weight-shifting), with data compared against a Bertec force platform.
- Statistical analyses included intraclass correlation coefficients, Pearson's/Spearman's correlations, Bland-Altmann plots, and mean absolute error (MAE).
Main Results:
- Mechanical tests demonstrated excellent reliability and validity, with hysteresis and non-linearity below 0.2% full-scale.
- Dynamic trials showed excellent cross-correlation, though MAE was higher for sit-to-stand compared to weight-shifting, potentially due to the 10 Hz sampling rate.
- The modified scales achieved high accuracy, detecting force changes as small as 0.01 kg.
Conclusions:
- Digital bathroom scales can be affordably modified to yield accurate vGRF data.
- These modified scales present a cost-effective solution for assessing vGRF in clinical and research settings.
- Further investigation into optimizing sampling rates may enhance dynamic measurement accuracy.
More Related Videos
08:40Quantifying Arms and Legs Contributions during Repetitive Electrically-Assisted Sit-To-Stand Exercise in Paraplegics: A Pilot Study
Published on: November 11, 2022
14:52Computerized Dynamic Posturography for Postural Control Assessment in Patients with Intermittent Claudication
Published on: December 11, 2013