Related Experiment Video
Updated: Nov 25, 2025

Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
Percutaneous Axillary Access for Placement of Microaxial Ventricular Support Devices: The Axillary Access Registry to
James M McCabe1, Amir A Kaki2, Duane S Pinto3
1Division of Cardiology, University of Washington, Seattle (J.M.M., W.L.).
Background:
There has been increasing utilization of short-term mechanical circulatory support devices for a variety of clinical indications. Many patients have suboptimal iliofemoral access options or reasons why early mobilization is desirable. Axillary artery access is an option for these patients, but little is known about the utility of this approach to facilitate short-term use for circulatory support with microaxial pump devices.
Methods:
The Axillary Access Registry to Monitor Safety (ARMS) was a prospective, observational multicenter registry to study the feasibility and acute safety of mechanical circulatory support via percutaneous upper-extremity access.
Results:
One hundred and two patients were collected from 10 participating centers. Successful device implantation was 98% (100 of 102). Devices were implanted for a median of 2 days (interquartile range, 0-5 days; range, 0-35 days). Procedural complications included 10 bleeding events and 1 stroke. There were 3 patients with brachial plexus-related symptoms all consisting of C8 tingling and all arising after multiple days of support. Postprocedural access site hematoma or bleeding was noted in 9 patients. Device explantation utilized closure devices alone in 61%, stent grafts in 17%, balloon tamponade facilitated closure in 15%, and planned surgical explant in 5%. Duration of support appeared to be independently associated with a 1.1% increased odds of vascular complication per day ([95% CI, 0.0%-2.3%] P=0.05).
Conclusions:
Percutaneous axillary access for use with microaxial support pumps appears feasible with acceptable rates of bleeding despite early experience. Larger studies are necessary to confirm the pilot data presented here.
More Related Videos
Related Concept Videos
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization II: Right Heart Catheterization
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.