Patient arm position during quantitative bone single-photon emission computed tomography/computed tomography acquisition can affect image quality and quantitative accuracy: a phantom study
Related Experiment Video
Updated: Nov 25, 2025

Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Noriaki Miyaji1,2, Kenta Miwa3, Kazuki Motegi1
1Department of Nuclear Medicine, Cancer Institute Hospital of Japanese Foundation for Cancer Research, Tokyo.
Purpose:
The present study used a phantom to determine the effects of various arm positions on bone SPECT/computed tomography (CT) images and the optimal arm position to acquire good-quality and quantitatively accurate images.
Materials And Methods:
We designed a phantom study of five simulated arm positions that are assumed during SPECT image acquisition. All SPECT data were acquired during a total of 120 projections of 10 and 100 s/view over 360° in a non-circular mode and reconstructed using Flash 3D (Siemens Healthineers). We evaluated contrast (QH,17 mm), image noise (NB,17 mm), contrast-to-noise ratios (QNRs), and visual scores according to the guidelines for bone SPECT acquisition protocols published by the Japanese Society of Nuclear Medicine Technology. The SUVmean, SUVmax, and SUVpeak were calculated and quantitative errors were evaluated using the recovery coefficient (RC) and the root means square error (RMSE).
Results:
The spatial resolution of SPECT images was better when the arms were down than raised with simulated shoulder disorders. Raised arms with shoulder disorders significantly increased the NB,17 mm and decreased the QH,17 mm, and the QNR in each image differed over a range from 2.2 to 5.2. The visual score was >1.5 with the arms down, raised normally, and raised with moderate shoulder disorders. The SUVmax and SUVpeak were overestimated compared with 100-min data for all images, whereas SUVmean was underestimated. Raised arms with a shoulder disorder decreased RCmax, and RCmean and RCpeak suppressed differences among arm positions. In addition, RMSE with the arms down and raised normally were close to that for 100-min data.
Conclusion:
Bone SPECT images with good quality and quantitative accuracy can be acquired with patients holding their arms down by their sides. This will help patients with shoulder pain who have difficulties raising their arms.
More Related Videos
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
08:04Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...