Mammography Diagnostic Reference Levels (DRLs) in Ghana
E Dzidzornu1, S K Angmorterh2, B B Ofori-Manteaw3
1Department of Medical Imaging, School of Allied Health Sciences, University of Health and Allied Sciences (UHAS), Ho, Ghana. Electronic address: https://twitter.com/BettyManteaw.
Introduction:
Diagnostic Reference Levels (DRLs) are essential for optimisation in mammography. A local DRL for screen-film mammography has been established in Ghana but none exists for the digital mammography systems. Furthermore, technological advancement is phasing out the use of screen-film mammography and replacing it with digital mammography systems. This study aims to establish the local DRLs used in digital mammography across three institutions in Ghana to guide mammography practice.
Methods:
Average glandular dose (AGD), compressed breast thickness (CBT), age of patients, entrance surface exposure (ESE), kVp, and mAs were retrospectively extracted from three digital mammography systems. The 75th and 95th percentile values were obtained for the AGD of each mammography projection and at CBT of 60 ± 5 mm. The correlation between the AGD and CBT, kVp, mAs, and ESE were investigated.
Results:
The 75th percentile for the AGD at CBT of 60 ± 5 mm for Centres 1, 2, 3, and all centres were 2.3, 1.8, 2.1, and 2.0 mGy respectively. The DRLs obtained were comparably higher than international studies except those of the United Kingdom. The AGD showed a strong positive correlation with the CBT, kVp, mAs, and ESE. There was variability in the AGD applied across the three centres for the craniocaudal (CC) and mediolateral oblique (MLO) projections. The mean AGD, mAs, and ESE for all the three centres and per centre recorded were higher than previous studies, but the mean kVp and CBT were lower than previous studies.
Conclusion:
The higher DRLs estimated in this preliminary study indicates that there is a need for dose optimisation in digital mammography practice in Ghana to improve radiation protection.
Implications For Practice:
The findings will guide the process of optimisation and limit the variations in the radiation dose during mammography practice.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...