Ensuring both velocity and spatial responses robust to field inhomogeneities for velocity-selective arterial spin
Dapeng Liu1,2, Wenbo Li1,2, Feng Xu1,2
1Department of Radiology, Johns Hopkins University School of Medicine, Baltimore, MD, USA.
Purpose:
To evaluate both velocity and spatial responses of velocity-selective arterial spin labeling (VS-ASL), using velocity-insensitive and velocity-compensated waveforms for control modules, as well as a novel dynamic phase-cycling approach, at different B0 / field inhomogeneities.
Methods:
In the presence of imperfect refocusing, the mechanism of phase-cycling the refocusing pulses through four dynamics was first theoretically analyzed with the conventional velocity-selective saturation (VSS) pulse train. Numerical simulations were then deployed to compare the performance of the Fourier-transform based velocity-selective inversion (FT-VSI) with these three different schemes in terms of both velocity and spatial responses under various B0 / conditions. Phantom and human brain scans were performed to evaluate the three methods at scales of 0.8, 1.0, and 1.2.
Results:
The simulations of FT-VSI showed that, under nonuniform B0 / conditions, the scheme with velocity-insensitive control was susceptible to DC bias of the static spins as systematic error, while the scheme with velocity-compensated control had deteriorated velocity-selective labeling profiles and, thus, reduced labeling efficiency. Through numerical simulation, phantom scans, and brain perfusion measurements, the dynamic phase-cycling method demonstrated considerable improvements over these issues.
Conclusion: