Toward real-time verification for MLC tracking treatments using time-resolved EPID imaging
Benjamin J Zwan1,2, Vincent Caillet2,3, Jeremy T Booth2,4
1School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW, Australia.
This study developed a real-time system using an electronic portal imaging device (EPID) to verify the geometric accuracy of multileaf collimator (MLC) tracking during radiotherapy. The system successfully detected delivery errors, ensuring precise tumor targeting during treatment.
Area of Science:
- Medical Physics
- Radiation Oncology
- Image Analysis
Background:
- Multileaf collimator (MLC) tracking in radiotherapy adapts to intrafraction tumor motion, introducing risks of delivery errors undetectable by traditional methods.
- Real-time verification is crucial for ensuring accuracy in adaptive radiotherapy techniques like MLC tracking.
Purpose of the Study:
- To develop and validate a system for real-time geometric verification of MLC tracking treatments using an electronic portal imaging device (EPID).
Main Methods:
- Utilized volumetric modulated arc therapy (VMAT) with MLC tracking, acquiring images at 9.57 fps with an EPID.
- Extracted MLC positions from image frames to assess field location, size, and shape accuracy.
- Validated the technique through phantom studies and comparison with treatment log files, then applied to clinical lung SABR patient data.
Main Results:
- Phantom validation detected field location errors with sub-millimeter uncertainty (0.28-0.38 mm).
- Clinical lung SABR deliveries showed field location errors of 1.6 mm (parallel) and 4.9 mm (perpendicular) to MLC motion.
- No correlation was found between geometric errors and dosimetric errors in the target volume for the patient cohort.
Conclusions:
- A novel system for real-time delivery verification of MLC tracking was successfully developed using time-resolved EPID imaging.
- The system independently verifies the geometric accuracy of the MLC during radiotherapy, enhancing treatment safety.
- This technique provides a critical tool for quality assurance in advanced radiotherapy delivery.