Related Experiment Video
Updated: Jul 31, 2026

Fabrication of Carbon Nanotube High-Frequency Nanoelectronic Biosensor for Sensing in High Ionic Strength Solutions
Published on: July 22, 2013
Use of biocompatible redox-active polymers based on carbon nanotubes and modified organic matrices for development of
V A Arlyapov1, A S Kharkova1, S K Kurbanaliyeva1
1Tula State University, 92 Lenin Pr., Tula, 300012, Russia.
Abstract:
This work investigated the use of redox-active polymers based on bovine serum albumin and chitosan, covalently bound to mediators neutral red and ferrocene and containing carbon nanotubes, for immobilization of Paracoccus yeei VKM B-3302 bacteria. The structures of produced polymers were studied by IR spectroscopy and scanning electron microscopy. Cyclic voltammetry and impedance spectroscopy found the electrochemical characteristics of the investigated systems: the heterogeneous electron transfer rate constant, the constant of the rate of interaction with P. yeei bacteria and the impedance. The systems containing carbon nanotubes and ferrocene-based redox-active polymer proved to be the most promising. Biosensors formed using the hybrid polymers had a high sensitivity with the lower boundary of 0.1 mg/dm3 of the detected BOD5 concentrations and a high correlation (R = 0.9916) with the standard BOD assay of surface water samples.