Related Experiment Video
Updated: Nov 23, 2025

09:04
A Simple Bioassay for the Evaluation of Vascular Endothelial Growth Factors
Published on: March 15, 2016
10.0K
An Automated Assay for Growth Differentiation Factor 15.
Kai C Wollert1, Tibor Kempf1, Evangelos Giannitsis2
1Division of Molecular and Translational Cardiology, Department of Cardiology and Angiology, Hannover Medical School, Hannover, Germany.
The Journal of Applied Laboratory Medicine
|December 31, 2020
Summary
The new Elecsys GDF-15 assay shows strong analytical performance for measuring Growth Differentiation Factor 15 (GDF-15) in routine lab conditions. This automated immunoassay offers reliable GDF-15 quantification in both serum and plasma samples.
Area of Science:
- Clinical Chemistry
- Biomarker Discovery
- Cardiovascular Disease Research
Background:
- Growth Differentiation Factor 15 (GDF-15) is a key biomarker for cardiovascular disease (CVD) burden and risk.
- Accurate and efficient measurement of GDF-15 is crucial for clinical applications.
Purpose of the Study:
- To evaluate the analytical performance of a new, fully automated electrochemiluminescence immunoassay (Elecsys® GDF-15 assay) for GDF-15 measurement.
- To assess the assay's precision, reproducibility, and correlation with existing methods.
Main Methods:
- Six laboratories assessed the Elecsys® GDF-15 assay under routine conditions.
- Evaluated within-run, within-laboratory, and between-laboratory precision.
- Compared Elecsys assay results with a research immunoradiometric assay (IRMA) and ELISA, and measured GDF-15 in 739 healthy individuals.
Main Results:
- Excellent precision with CVs ranging from 0.7% to 8.6% for within-run and within-laboratory measures.
- High between-laboratory precision (CVs 7.1%–8.9%) and comparable GDF-15 recovery across sample types and reagent lots.
- Strong correlation and agreement between the Elecsys assay and IRMA/ELISA in the clinically relevant range.
Conclusions:
- The Elecsys GDF-15 assay exhibits robust analytical performance in routine laboratory settings.
- This automated immunoassay provides a reliable method for quantifying GDF-15 in both serum and plasma.
- Findings support the use of this assay for GDF-15 assessment in cardiovascular research and clinical practice.