Comparative study between fixed-angle and polyaxial screws in distal radius fixation with two volar locking plates
Álvaro Toro-Aguilera1, Pablo Martínez-Galarza2, Pilar Camacho-Carrasco3
1Department of Orthopaedics and Trauma Surgery, Hospital General de Granollers, Avinguda Francesc Ribas s/n, 08402 Granollers, Spain; Department of Orthopaedics and Trauma Surgery, Hospital de la Santa Creu i Sant Pau, Universitat Autònoma de Barcelona, c/Sant Antoni Maria Claret, 167, 08025 Barcelona, Spain.
Fixed-angle and polyaxial screws show similar protrusion rates in distal radius fractures. Polyaxial screws better preserve radial height, while fixed-angle screws may have shorter protrusions. Dorsal screw protrusion is common with both systems.
Area of Science:
- Orthopedic surgery
- Biomedical engineering
- Radiology
Background:
- Distal radius fractures are common, often treated with volar locking plates (VLPs).
- Numerous plate-screw designs exist, but no single system demonstrates clear superiority.
- Understanding screw behavior is crucial for optimal fracture fixation and patient outcomes.
Purpose of the Study:
- To compare the efficacy of fixed-angle (FA) versus polyaxial (PA) screws in distal radius VLP fixation.
- To evaluate screw protrusion, reduction preservation, and intra-articular penetration between the two screw types.
Main Methods:
- Prospective cohort study of 80 patients with distal radius fractures treated with VLPs.
- 40 patients received PA VLPs, and 40 received FA VLPs via the flexor carpi radialis approach.
- Postoperative CT scans assessed screw protrusion; loss of reduction was measured over 12 months.
Main Results:
- No significant difference in the rate of dorsal screw protrusion between PA and FA groups.
- Fixed-angle screws resulted in shorter protruding screws compared to polyaxial screws (p<0.05).
- Polyaxial screws demonstrated better preservation of radial height (p<0.05), while loss of reduction was similar between groups.
Conclusions:
- Both FA and PA screw systems exhibit comparable screw protrusion rates in distal radius VLP fixation.
- Polyaxial screws offer an advantage in maintaining radial height, whereas fixed-angle screws may be associated with shorter dorsal screw penetration.
- Dorsal and intra-articular screw protrusion are more frequent than anticipated, regardless of screw type.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Related Concept Videos
Screw: Problem Solving
To evaluate the...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.