Virucidal efficacy of glutaraldehyde for instrument disinfection
Florian H H Brill1, Britta Becker1, Daniel Todt2
1Dr. Brill + Partner GmbH Institute for Hygiene and Microbiology, Bremen, Germany.
Abstract:
Aim: Glutaraldehyde (GDA) is an active ingredient in many instrument disinfectants and is effective against a broad spectrum of microorganisms. In the past, the virus-inactivating properties of these products were mainly claimed based on quantitative suspension tests with different test viruses. Recently, however, a European Norm EN 17111:2018 has been published which allows examination of instrument disinfectants in a surface carrier test, simulating practical conditions. Therefore, it is of interest to evaluate GDA for the ability to inactivate the viruses used in this European Norm as test viruses. Methods: The virucidal efficacy of GDA as the active ingredient in instrument disinfectants was evaluated with 4 different test viruses in a method simulating practical conditions (EN 17111:2018). Results: With a fixed exposure time of five minutes at 20°C, 100 ppm GDA were necessary to inactivate vaccinia virus, classifying it as a limited spectrum virucidal activity for pre-cleaning products. For adenovirus, 125 ppm GDA were required, whereas for murine norovirus as a surrogate for human norovirus, 4,000 ppm GDA were required for a significant reduction of viral titres. Both non-enveloped viruses must be tested to prove virucidal activity in EN 17111:2018. But even 4,000 ppm were not enough to yield a 4 log10 reduction of the murine parvovirus at 20°C. This virus is only required as a test virus using this method if temperatures ≥40°C are used. Conclusion: GDA, as the active ingredient of many instrument disinfectants, shows virucidal efficacy at 20°C. The necessary concentrations are strongly dependent on the stability of the test viruses. Due to the high stability of murine norovirus, GDA levels of 4,000 ppm were required to inactivate this virus within the 5-minute exposure time.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Chemical Agents for Microbial Control
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Hand hygiene
Hand washing...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration