Related Experiment Video
Updated: Nov 23, 2025

A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
Automated control of fraction of inspired oxygen: is it time for widespread adoption?
Souvik Mitra1,2, Douglas McMillan1
1Division of Neonatal Perinatal Medicine, Department of Pediatrics, Dalhousie University and IWK Health.
Purpose Of Review:
Over the past two decades, numerous algorithms for automated control of the fraction of inspired oxygen (FiO2) have been developed and incorporated into contemporary neonatal ventilators and high-flow devices in an attempt to optimize supplemental oxygen therapy in preterm infants. This review explores whether current evidence is sufficient to recommend widespread adoption of automated oxygen control in neonatal care.
Recent Findings:
To date, 15 studies have compared automated versus manual control of FiO2 in preterm infants on respiratory support. This includes four new randomized cross-over trials published in the last 2 years. Available evidence consistently demonstrates a significant improvement in time spent within the target saturation range with automated FiO2 control. There are fewer episodes of severe hypoxemia and fewer manual FiO2 adjustments with automated oxygen control. Nursing workload may be reduced. However, no currently completed studies report on clinical outcomes, such as chronic lung disease or retinopathy of prematurity.
Summary:
Automated oxygen control appears to be a reasonable option for FiO2 titration in preterm infants on respiratory support, if resources are available, and might substantially reduce nursing workload. Further randomized clinical trials to explore its effects on clinical outcomes are required.
Related Concept Videos
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...