An electronic portal image device (EPID)-based multiplatform rapid daily LINAC QA tool
Yangguang Ma1, Xuemin Wang2, Rizhen Mai3
1Department of Radiation Oncology, The First Affiliated Hospital of Zhengzhou University, Zhengzhou, China.
Journal of Applied Clinical Medical Physics
|January 7, 2021
Summary
A new daily quality research tool (DQRT) using an Electronic Portal Image Device (EPID) efficiently checks linear accelerators (LINACs) for both flattening filter (FF) and flattening filter-free (FFF) photon beams, offering an economic alternative for quality assurance.
Area of Science:
- Medical Physics
- Radiation Oncology
- Quality Assurance
Background:
- Daily quality assurance (QA) is critical for multiplatform linear accelerators (LINACs).
- Existing QA tools may be time-consuming or costly, especially for institutions with limited resources.
- The need for an efficient and economical daily quality research tool (DQRT) for LINACs is evident.
Purpose of the Study:
- To develop an efficient and economic daily quality research tool (DQRT) for LINACs.
- To utilize an Electronic Portal Image Device (EPID) for daily checks of both flattening filter (FF) and flattening filter-free (FFF) photon beams.
- To provide a cost-effective QA solution for radiation therapy.
Main Methods:
- EPID calibration followed by analysis of monitored parameters from open and wedge portal images.
- Verification of EPID position accuracy, long-term stability, and linearity between image gray value and exposure.
- Evaluation of output and beam quality stability for 6-MV FF and FFF beams, including introduced setup errors, and comparison with commercial QA tools (FC65-G, Matrixx).
Main Results:
- All monitored parameters demonstrated excellent repeatability.
- A strong linear correlation was observed between EPID image grayscale values and exposure.
- The DQRT showed negligible differences compared to FC65-G and Matrixx, with consistent results over a 2-month evaluation period on three LINACs.
Conclusions:
- The developed DQRT is a time-efficient and effective tool for daily LINAC QA.
- It provides QA results comparable or superior to traditional methods, with richer data.
- The DQRT presents a viable, economical alternative for hospitals with limited QA time or budget constraints.