Related Experiment Video
Updated: Nov 22, 2025

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
CT Fluoroscopy for Image-Guided Procedures: Physician Radiation Dose During Full-Rotation and Partial-Angle CT
Emily A Knott1, Sean D Rose2, Martin G Wagner2
1Department of Radiology, University of Wisconsin, 1111 Highland Ave, Madison, WI, 53705.
Purpose:
To determine physician radiation exposure when using partial-angle computed tomography (CT) fluoroscopy (PACT) vs conventional full-rotation CT and whether there is an optimal tube/detector position at which physician dose is minimized.
Materials And Methods:
Physician radiation dose (entrance air kerma) was measured for full-rotation CT (360°) and PACT (240°) at all tube/detector positions using a human-mimicking phantom placed in a 64-channel multidetector CT. Parameters included 120 kV, 20- and 40-mm collimation, and 100 mA. The mean, standard deviation, and increase/decrease in physician dose compared with a full-rotation scan were reported.
Results:
Physician radiation exposure during CT fluoroscopy with PACT was highly dependent on the position of the tube/detector during scanning. The lowest PACT physician dose was when the physician was on the detector side (center view angle 116°; -35% decreased dose vs full-angle CT). The highest PACT physician dose was with the physician on the tube side (center view angle 298°; +34% increased dose vs full-angle CT), all doses P <.05 vs full-rotation CT.
Conclusions:
Partial-angle CT has the potential to both significantly increase or decrease physician radiation dose during CT fluoroscopy-guided procedures. The detector/tube position has a profound effect on physician dose. The lowest dose during PACT was achieved when the physician was located on the detector side (ie, distant from the tube). This data could be used to optimize CT fluoroscopy parameters to reduce physician radiation exposure for PACT-capable scanners.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...