Influence of voxel size on cone beam computed tomography artifacts arising from the exomass
Amanda P Candemil1, Benjamin Salmon2, Gláucia M B Ambrosano3
1Researcher, Division of Oral Radiology, Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas, Piracicaba-SP, Brazil.
Objectives:
The objective of this study was to evaluate the influence of voxel size on artifacts arising from the exomass in cone beam computed tomography (CBCT).
Study Design:
An imaging phantom was scanned using 2 CBCT units, each adjusted to 2 voxel sizes: 0.2 and 0.3 mm. From 1 to 3 metal inserts of titanium, cobalt-chromium, or amalgam were placed in the exomass and additional CBCT scans were acquired. Mean voxel gray values were obtained from 16 homogeneous areas of the phantom and averaged, and the standard deviation was calculated to obtain voxel gray value variability. The data were analyzed using 2-way analysis of variance, Tukey, and Dunnett tests (α =.05).
Results:
Overall, mean voxel gray values and voxel gray value variability did not differ significantly between CBCT scans obtained with voxel sizes of 0.2 and 0.3 mm for either CBCT model tested (P > .05). Despite some exceptions in which significant differences were observed between the 2 voxel sizes (P < .05), the mean voxel gray values and voxel gray value variability resulting from different metal compositions and, in most situations, for different numbers of metal inserts in the exomass were not affected.
Conclusion:
Voxel size has little influence on exomass-related CBCT artifacts.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...