Related Experiment Video
Updated: Nov 21, 2025

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Regional lung function measures determined by electrical impedance tomography during repetitive ventilation
L Lasarow1, B Vogt1, Z Zhao2,3
1Department of Anaesthesiology and Intensive Care Medicine, University Medical Centre Schleswig-Holstein, Campus Kiel, Kiel, Germany.
Objective:
Current standards for conducting spirometry examinations recommend that the ventilation manoeuvres needed in pulmonary function testing are carried out repeatedly during sessions. Chest electrical impedance tomography (EIT) can determine the presence of ventilation heterogeneity during such manoeuvres, which increases the information content derived from such examinations. The aim of this study was to characterise regional lung function in patients with chronic obstructive pulmonary disease (COPD) during repetitive forced full ventilation manoeuvres. Regional lung function measures derived from these manoeuvres were compared with quiet tidal breathing.
Approach:
Sixty hospitalised patients were examined during up to three repeated ventilation manoeuvres. Acceptable spirometry manoeuvres were performed and EIT recordings suitable for analysis obtained in 53 patients (12 women, 41 men; age: 68 ± 12 years (mean ± SD)). Pixel values of tidal volume, forced full inspiratory and expiratory volume in 1 s, and forced inspiratory and expiratory vital capacity were calculated from the EIT data. Spatial ventilation heterogeneity was assessed using the coefficient of variation, global inhomogeneity index, and centres and regional fractions of ventilation. Temporal inhomogeneity was determined by examining the pixel expiration times needed to exhale 50% and 75% of regional forced vital capacity.
Main Results:
All EIT-derived measures of regional lung function showed reproducible results during repetitive examinations. Parameters of spatial heterogeneity obtained from quiet tidal breathing were comparable with the measures derived from the forced manoeuvres.
Significance:
Measures of spatial and temporal ventilation heterogeneity obtained in COPD patients by EIT provide comparable findings during repeated examinations within one testing session. Quiet tidal breathing generates similar information on ventilation heterogeneity as forced manoeuvres that demand a high amount of patient effort.
More Related Videos
08:44Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
Published on: February 2, 2024
05:56Author Spotlight: Enhancing Diagnostic Strategies and Biomarker Development for Comprehensive Lung Function Analysis
Published on: August 9, 2024
Related Concept Videos
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: