A pilot study for intraocular pressure measurements based on vibroacoustic parameters
Deukha Kim1, Youngbeen Chung1, Yeji Yeon2
1Department of Mechanical Engineering, Hanyang University College of Engineering, Seoul, Korea.
Scientific Reports
|January 14, 2021
Summary
This study found that eyeball vibration properties, specifically the change in magnitude of vibration response (CMVR), correlate with intraocular pressure (IOP). This suggests CMVR could be a new method for continuous IOP measurement.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Vibroacoustics
Background:
- Intraocular pressure (IOP) is a critical factor in diagnosing and managing glaucoma.
- Current methods for IOP measurement can be invasive or lack continuous monitoring capabilities.
Purpose of the Study:
- To identify vibroacoustic properties linked to intraocular pressure (IOP) variations.
- To propose a novel method for IOP measurement based on these vibroacoustic properties.
Main Methods:
- Ten ex vivo porcine eyeballs were utilized.
- Vibration response was measured using an accelerometer attached to the eyeball.
- Intraocular pressure was manipulated by injecting balanced salt solution and measured with a tonometer.
Main Results:
- Resonance frequency and the change in the magnitude of the vibration response (CMVR) increased with elevated IOP.
- CMVR demonstrated a strong correlation with IOP (p-value < 0.0001).
- A linear mixed effects model confirmed the relationship between vibroacoustic properties and IOP.
Conclusions:
- Vibroacoustic properties of the eyeball are significantly correlated with changes in intraocular pressure.
- The CMVR shows potential as a novel parameter for IOP assessment.
- This research paves the way for future continuous IOP monitoring systems.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
2.0K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
2.0K
Measurement of Fluid Pressure
419
Fluid pressure is commonly measured using devices called manometers, which rely on liquid columns to indicate pressure differences. The height of a liquid column in a manometer reflects the pressure exerted by the fluid, providing a simple yet effective means of measurement. Different types of manometers serve specific purposes based on their configurations and the type of fluids involved.
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
419
Measurement of Blood Pressure
2.2K
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
2.2K