Related Experiment Video
Updated: Nov 21, 2025

Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
Published on: October 7, 2015
Assessing the Validity of Aural Thermometry for Measuring Internal Temperature in Patients With Exertional Heat
Margaret C Morrissey1, Samantha E Scarneo-Miller1,2, Gabrielle E W Giersch1
1Korey Stringer Institute, Department of Kinesiology, University of Connecticut, Storrs.
Context:
The use of aural thermometry as a method for accurately measuring internal temperature has been questioned. No researchers have examined whether aural thermometry can accurately measure internal body temperature in patients with exertional heat stroke (EHS).
Objective:
To examine the effectiveness of aural thermometry as an alternative to the criterion standard of rectal thermometry in patients with and those without EHS.
Design:
Cross-sectional study.
Setting:
An 11.3-km road race.
Patients Or Other Participants:
A total of 49 patients with EHS (15 men [age = 38 ± 17 years], 11 women [age = 28 ± 10 years]) and 23 individuals without EHS (10 men [age = 62 ± 17 years], 13 women [age = 45 ± 14 years]) who were triaged to the finish-line medical tent for suspected EHS.
Main Outcome Measure(S):
Rectal and aural temperatures were obtained on arrival at the medical tent for patients with and those without EHS and at 8.3 ± 5.2 minutes into EHS treatment (cold-water immersion) for patients with EHS.
Results:
The mean difference between temperatures measured using rectal and aural thermometers in patients with EHS at medical tent admission was 2.4°C ± 0.96°C (4.3°F ± 1.7°F; mean rectal temperature = 41.1°C ± 0.8°C [106.1°F ± 1.4°F]; mean aural temperature = 38.8°C ± 1.1°C [101.8°F ± 2.0°F]). Rectal and aural temperatures during cold-water immersion in patients with EHS were 40.4°C ± 1.0°C (104.6°F ± 1.8°F) and 38.0°C ± 1.2°C (100.3°F ± 2.2°F), respectively. Rectal and aural temperatures for patients without EHS at medical tent admission were 38.8°C ± 0.87°C (101.9°F ± 1.6°F) and 37.2°C ± 1.0°C (99.1°F ± 1.8°F), respectively.
Conclusions:
Aural thermometry is not an accurate method of diagnosing EHS and should not be used as an alternative to rectal thermometry. Using aural thermometry to diagnosis EHS can result in catastrophic outcomes, such as long-term sequelae or fatality.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...