Safety and image quality of MR-conditional external fixators for 1.5 Tesla extremity MR
David H Ballard1, John D Garrett2, Alberto A Simoncini2
1Mallinckrodt Institute of Radiology, Washington University School of Medicine, 510 S Kings Highway Blvd Campus Box 8131, St Louis, MO, 63110, USA. davidballard@wustl.edu.
Purpose:
To evaluate the safety and image quality of extremity MR examinations performed with two MR conditional external fixators located in the MR bore.
Materials And Methods:
Single-center retrospective study of a prospectively maintained imaging dataset that evaluated MR examinations of extremities in patients managed with external fixations instrumentation and imaged on a single 1.5T MR scanner. The fixation device was one of two MR-conditional instrumentation systems: DuPuy Synthes (aluminum, stainless steel, carbonium and Kevlar) or Dolphix temporary fixation system (PEEK-CA30). Safety events were recorded by the performing MR radiologic technologist. A study musculoskeletal radiologist assessed all sequences to evaluate for image quality, signal- and contrast-to-noise ratios (SNR/CNR), and injury patterns/findings.
Results:
In the 13 men and 9 women with a mean age of 42 years (range 18 to 72 years), most patients (19/22 patients; 86%) were involved with trauma resulting in extremity injury requiring external fixation. MR examinations included 19 knee, 2 ankle, and 1 elbow examinations. There were no adverse safety events, heating that caused patient discomfort, fixation dislodgement/perturbment, or early termination of MR examinations. All examinations were of diagnostic quality. Fat-suppressed proton density sequences had significantly higher SNR and CNR compared to STIR (p = 0.01 to 0.04). The lower SNR of STIR and increased quality of fat-suppressed proton density during the study period led to the STIR sequence being dropped in standard MR protocol.
Conclusion:
MR of the extremity using the two study MR conditional external fixators within the MR bore is safe and feasible.
More Related Videos
08:55Use of a Foot-Induced Digitally Controlled Resistance Device for Functional Magnetic Resonance Imaging Evaluation in Patients with Foot Paresis
Published on: July 7, 2023
10:09Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI