Related Experiment Video
Updated: Nov 21, 2025

Agarose-based Tissue Mimicking Optical Phantoms for Diffuse Reflectance Spectroscopy
Published on: August 22, 2018
Water equivalence of a solid phantom material for radiation dosimetry applications
Maegan A Gargett1, Adam R Briggs1, Jeremy T Booth1,2
1Northern Sydney Cancer Centre, Royal North Shore Hospital, Sydney, NSW, Australia.
Abstract:
Radiological water equivalence of solid phantoms used for radiotherapy is often desired, but is non-trivial to achieve across the range of therapeutic energies. This study evaluated the water equivalence of a new solid phantom material in beam qualities relevant to radiotherapy applications. In-phantom measured depth distributions were compared to that in water to assess the relative attenuation and scatter characteristics of the material. The phantom material was found to be dosimetrically equivalent to water within (1.0 ± 1.0)% for megavoltage photon beam qualities, (1.5 ± 1.3)% for megavoltage electron beam qualities, (1.5 ± 1.5)% for medium-energy kilovoltage X-rays and (3.0 ± 1.5)% for low-energy kilovoltage X-rays.