Related Experiment Video
Updated: Nov 21, 2025

Treating Low Back Pain in Failed Back Surgery Patients with Multicolumn-lead Spinal Cord Stimulation
Published on: June 26, 2018
Unexpected extrusion of the implantable pulse generator of the spinal cord stimulator - A case report
Eun-Ji Choi1,2, Hyun-Su Ri1,2, Hyeonsoo Park1
1Department of Anesthesia and Pain Medicine, Pusan National University Yangsan Hospital, Pusan National University School of Medicine, Yangsan, Korea.
Background:
Despite significant technological advances in the implantable pulse generator (IPG), complications can still occur. We report a case that unexpected extrusion of the IPG of spinal cord stimulation (SCS) was promptly identified and successfully removed without any complications.
Case:
After a car accident 4 years ago, a 55-year-old male who was diagnosed with complex local pain syndrome in his right leg. The SCS was inserted with 2 leads, with the IPG being implanted in the right lower abdomen region. Four years later, he developed extrusion of the IPG from his abdominal region. This unexpected extrusion may have been related to pressure necrosis caused by continued compression of pocket site where a belt was frequently tied. The IPG and the leads were successfully removed without infection occurring.
Conclusions:
To prevent unexpected extrusion of IPG, it is necessary to consider in advance whether the pocket site is pressed against the belt.