Automation of routine coagulation testing using a random access centrifugal analyzer.

J A Rowell1, M J Elms, J Young

  • 1Division of Hematology, Royal Brisbane Hospital, Queensland, Australia.

American Journal of Clinical Pathology
|March 1, 1988
PubMed
Summary

Automating hemostasis testing with a centrifugal analyzer significantly reduced workload and costs. This system accurately performed routine coagulation tests, improving lupus anticoagulant detection and offering flexibility for diverse pathology assays.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Cell Suspensions from Collagenase Digestion of Bone Marrow Trephine Biopsy Specimens-its Use in Lymphoma.

Leukemia & lymphoma·2016
Same author

Multidimensional Scaling: A Possible Technique for Examining Male and Female Occupational Perceptions and Preferences.

Multivariate behavioral research·2016
Same author

Enhancing haemophilia care through registries.

Haemophilia : the official journal of the World Federation of Hemophilia·2014
Same author

Enzyme replacement therapy for Gaucher disease in Australia.

Internal medicine journal·2005
Same author

Early diagnosis of lymphedema using multiple frequency bioimpedance.

Lymphology·2001
Same author

Therapeutic adjuncts for immediate transfer to the catheterization laboratory in patients with acute coronary syndromes.

The American journal of cardiology·2001

Area of Science:

  • Clinical Chemistry
  • Hematology
  • Laboratory Automation

Background:

  • Hemostasis laboratories manage routine coagulation testing workloads.
  • Accurate and efficient testing is crucial for patient diagnosis and management.

Purpose of the Study:

  • To evaluate a random-access centrifugal analyzer for routine hemostasis laboratory management.
  • To assess the analyzer's performance against manual methods for key coagulation tests.

Main Methods:

  • A centrifugal analyzer linked to a personal computer was used for prothrombin time (PT), activated partial thromboplastin time (APTT), thrombin clotting time (TCT), and fibrinogen (Fib) assays.
  • Tests were performed on 929 samples, including those from patients on anticoagulants and requiring coagulation screens.
  • Results were correlated with parallel manual tilt-tube techniques.

Related Experiment Videos

Main Results:

  • High correlation coefficients were observed for PT (0.99), APTT (0.96), and Fib (0.97).
  • TCT showed a lower correlation (0.72) due to manual test variability and sample-specific factors.
  • The analyzer accurately detected abnormalities, improved lupus anticoagulant detection, and reduced workload and reagent costs.

Conclusions:

  • A programmed centrifugal analyzer provides a safe, efficient, and flexible method for automating routine coagulation tests.
  • This automation enhances the diagnostic capabilities of hemostasis laboratories and offers potential for broader pathology applications.