Validation of a Portable Blood Gas Analyzer for Use in Challenging Field Conditions at High Altitude
Janek Nawrocki1, Michael Furian1, Aline Buergin1
1Department of Respiratory Medicine, Sleep Disorders Center, University Hospital of Zürich, Zurich, Switzerland.
Background:
Novel, portable blood gas analyzers (BGAs) may serve as essential point-of-care tools in remote regions, during air travel or in ambulance services but they have not been extensively validated.
Research Question:
We compared accuracy of a portable BGA to a validated stationary device.
Methods:
In healthy individuals and patients with chronic obstructive pulmonary disease participating in clinical field studies at different altitudes, arterial blood samples were obtained at rest and during exercise in a hospital at 760 m and in a high altitude clinic at 3100 m. Paired measurements by a portable BGA (EPOC, Siemens Healthcare) and a stationary BGA (Rapidpoint500, Siemens Healthcare) were performed to compute bias (mean difference) and limits of agreement (95% CI of bias).
Results:
Of 105 individuals, 248 arterial blood samples were analyzed, 108 at 760 m, 140 at 3100 m. Ranges of values measured by portable BGA were: pH 7.241-7.473, PaCO2 21.5-52.5 mmHg, and PaO2 45.5-107.1 mmHg. Bias (95% CI) between devices were: pH 0.007 (-0.029 to 0.044), PaCO2 -0.3 mmHg (-4.8 to 4.2), and PaO2 -0.2 mmHg (-9.1 to 4.7). For pH, agreement between devices was improved by the equation to correct pH by portable BGA = -1.37 + pH × 1.19; bias after correction -0.007 (-0.023 to 0.009). The portable BGA was easily handled and worked reliably.
Interpretation:
Accuracy of blood gas analysis by the portable BGA in comparison to the reference BGA was adequate for clinical use. Because of portability and ease of handling, portable BGA are valuable diagnostic tools for use in everyday practice as well as under challenging field conditions.
More Related Videos
08:44Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
Published on: February 2, 2024
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Related Concept Videos
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...