Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
Comparison of procedural efficacy and biophysical parameters between two competing cryoballoon technologies for
Sing-Chien Yap1, Ante Anic2, Toni Breskovic2
1Department of Cardiology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, The Netherlands.
Introduction:
Recently a novel cryoballoon system (POLARx, Boston Scientific) became available for the treatment of atrial fibrillation. This cryoballoon is comparable with Arctic Front Advance Pro (AFA-Pro, Medtronic), however, it maintains a constant balloon pressure. We compared the procedural efficacy and biophysical characteristics of both systems.
Methods:
One hundred and ten consecutive patients who underwent first-time cryoballoon ablation (POLARx: n = 57; AFA-Pro: n = 53) were included in this prospective cohort study.
Results:
Acute isolation was achieved in 99.8% of all pulmonary veins (POLARx: 99.5% vs. AFA-Pro: 100%, p = 1.00). Total procedure time (81 vs. 67 min, p < .001) and balloon in body time (51 vs. 35 min, p < .001) were longer with POLARx. After a learning curve, these times were similar. Cryoablation with POLARx was associated with shorter time to balloon temperature -30°C (27 vs. 31 s, p < .001) and -40°C (32 vs. 54 s, p < .001), lower balloon nadir temperature (-55°C vs. -47°C, p < .001), and longer thawing time till 0°C (16 vs. 9 s, p < .001). There were no differences in time-to-isolation (TTI; POLARx: 45 s vs. AFA-Pro 43 s, p = .441), however, POLARx was associated with a lower balloon temperature at TTI (-46°C vs. -37°C, p < .001). Factors associated with acute isolation differed between groups. The incidence of phrenic nerve palsy was comparable (POLARx: 3.5% vs. AFA-Pro: 3.7%).
Conclusion:
The novel cryoballoon is comparable to AFA-Pro and requires only a short learning curve to get used to the slightly different handling. It was associated with faster cooling rates and lower balloon temperatures but TTI was similar to AFA-Pro.