Intraocular lens power calculation for plus and minus lenses in high myopia using partial coherence interferometry
Matthias Fuest1, Niklas Plange1, David Kuerten1
1Department of Ophthalmology, RWTH Aachen University, Pauwelsstrasse 30, 52074, Aachen, Germany.
Purpose:
We assessed the accuracy of lens power calculation in highly myopic patients implanting plus and minus intraocular lenses (IOL).
Methods:
We included 58 consecutive, myopic eyes with an axial length (AL) > 26.0 mm, undergoing phacoemulsification and IOL implantation following biometry using the IOLMaster 500. For lens power calculation, the Haigis formula was used in all cases. For comparison, refraction was back-calculated using the Barrett Universal II (Barrett), Holladay I, Hill-RBF (RBF) and SRK/T formulae.
Results:
The mean axial length was 30.17 ± 2.67 mm. Barrett (80%), Haigis (87%) and RBF (82%) showed comparable numbers of IOLs within 1 diopter (D) of target refraction. Visual acuity (BSCVA) improved (p < 0.001) from 0.60 ± 0.35 to 0.29 ± 0.29 logMAR (> 28-days postsurgery). The median absolute error (MedAE) of Barrett 0.49 D, Haigis 0.38, RBF 0.44 and SRK/T 0.44 did not differ. The MedAE of Haigis was significantly smaller than Holladay (0.75 D; p = 0.01). All median postoperative refractive errors (MedRE) differed significantly with the exception of Haigis to SRK/T (p = 0.6): Barrett - 0.33 D, Haigis 0.25, Holladay 0.63, RBF 0.04 and SRK/T 0.13. Barrett, Haigis, Holladay and RBF showed a tendency for higher MedAEs in their minus compared to plus IOLs, which only reached significance for SRK/T (p = 0.001). Barrett (p < 0.001) and RBF (p = 0.04) showed myopic, SRK/T (p = 002) a hyperopic shift in their minus IOLs.
Conclusions:
In highly myopic patients, the accuracies of Barrett, Haigis and RBF were comparable with a tendency for higher MedAEs in minus IOLs. Barrett and RBF showed myopic, SRK/T a hyperopic shift in their minus IOLs.
Related Concept Videos
Focusing of Light in the Eye
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...