In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
Spherical center axial hinge knee prosthesis causes lower contact stress on tibial insert and bushing compared with
Jing-Yu Zhang1, Hao-Ran Zhang2, Dong-Mu Tian3
1Tianjin Hospital, Tianjin, China; Second Hospital of Tangshan, Hebei, China.
Background:
Motion axial system may affect contact stress of hinge knee prosthesis. However, it is unclear which axial system provides the better biomechanical effect. Therefore, the aim of this study was to compare the contact stress and stress distribution on the tibial insert and the bushing of hinge knee prostheses with a biaxial (BA) system and a spherical center axial (SA) system during a gait cycle.
Methods:
Three-dimensional finite-element (FE) models of the prostheses with different motion systems were included. The comparisons between experimental tests and FE analyses were performed to verify the models. Dynamic implicit FE analyses were performed to investigate the peak contact stresses and stress distributions on the tibial insert and the bushing.
Results:
The peak contact stresses on the tibial insert and the bushing of the BA prosthesis were higher than those of the SA prosthesis during most gait cycles. The contact time on the bushing is short in the SA prosthesis. The stress distributions on the superior surface of the tibial insert in the BA prosthesis were at the posterior side, but of the SA prosthesis were not fixed.
Conclusion:
The SA prosthesis has a lower peak contact stress on tibial insert and bushing than the BA prosthesis; in addition, the SA prosthesis has a 'self-adjustment' mechanism which could disperse high stress on the tibial insert to decrease the risk of wear and damage. The comparison could help designers and surgeons to better understand the future design of rotating hinge knee prostheses which should be able to achieve multiaxial motion and complete weight bearing by the tibial condylar to transmit the axial force better.
Related Concept Videos
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Eccentric Axial Loading in a Plane of Symmetry
Pivot Bearings
A pivot bearing is a specialized type of bearing designed to support axial loads on a rotating shaft. The bearing surface, or the pivot, is positioned at the end of a shaft to support the axial thrust. The pivot may...
Bones of the Lower Limb: Femur and Patella
Bones of the Lower Limb: Tibia and Fibula
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...