Implications of dispersion in connecting capillaries for separation systems involving post-column flow splitting
Caden Gunnarson1, Thomas Lauer1, Harrison Willenbring1
1Gustavus Adolphus College, Department of Chemistry, St. Peter, MN, USA.
Abstract:
It is common practice in liquid chromatography to split the flow of the effluent exiting the analytical column into two or more parts, either to enable parallel detection (e.g., coupling the separation to two destructive detectors such as light scattering and mass spectrometry (MS)), or to accommodate flow rate limitations of a detector (e.g., electrospray ionization mass spectrometry). In these instances the user must make choices about split ratio and dimensions of connecting tubing that is used between the split point and the detector, however these details are frequently not mentioned in the literature, and rarely justified. In our own work we often split the effluent following the second dimension (2D) column in two-dimensional liquid chromatography systems coupled to MS detection, and we have frequently observed post 2D column peak broadening that is larger than we would expect to result from dispersion in the MS ionization source itself. For the present paper we describe a series of experiments aimed at understanding the impact of the split ratio and post-split connecting tubing dimensions on dispersion of peaks exiting an analytical column. We start with the simple idea - based on the principle of conservation of mass - that analyte peaks entering the split point are split into two parts such that the analyte mass (and thus peak volume) entering and exiting the split point is conserved, and directly related to the ratio of flow rates entering and exiting the split point. Measurements of peak width and variance after the split point show that this simple view of the splitting process - along with estimates of additional dispersion in the post-split tubing - is sufficient to predict peak variances at the detector with accuracy that is sufficient to guide experimental work (median error of about 10% over a wide range of conditions). We feel it is most impactful to recognize that flow splitting impacts apparent post-column dispersion not because anything unexpected happens in the splitting process, but because the split dramatically reduces the volume of the analyte peak, which then is more susceptible to dispersion in connecting tubing that would not cause significant dispersion under conditions where splitting is not implemented. These results will provide practitioners with a solid basis on which rational decisions about split ratios and dimensions of post-split tubing can be made.
Related Concept Videos
Capillary Electrophoresis: Instrumentation
Electrophoresis: Overview
There...
Principles Of Column Chromatography
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...