[Comparative analysis of the stress-strain state of standard and individual reconstructive plates]
1Donetsk National Medical University, Donetsk, Ukraine.
Objective:
The aim of the study was to analyze a three-dimensional finite element model of a post-resection defect of the lower jaw, design and producing of an individual reconstructive plate, compare the stress-strain of this plate using standard reconstructive «CONMET» plate for stabilizing the lower jaw segments through a simulation of the chewing load.
Material And Methods:
The study was carried out on two identical finite elementary models with bone defects in the body of the lower jaw by fixing with a standard reconstructive «CONMET» plate and designed reconstructive plate. In order to simulate the biomechanical loads on the plates when chewing, an impact of force during the occlusal contact of the incisor was applied. The magnitude of the vertical load was equal to 150 N. The condylar processes were fixed in all three directions to prevent reaction forces in the temporomandibular joint.
Results:
Comparative analysis of two fixation methods showed higher value of equivalent Von Mises stress in individual reconstructive plate than in standard plate.
Conclusion:
Individual reconstructive plate has better strength characteristics compared to the standard reconstructive «CONMET» plate.
More Related Videos
11:28A Coupled Experiment-finite Element Modeling Methodology for Assessing High Strain Rate Mechanical Response of Soft Biomaterials
Published on: May 18, 2015
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Related Concept Videos
Three-Dimensional Analysis of Strain
Stress Concentrations
Stress Concentrations
The stress...
True Stress and True Strain
In contrast, true stress offers a more precise portrayal. It is computed by dividing the...
Plastic Deformations
Plastic Deformations