Biomimetic phantom with anatomical accuracy for evaluating brain volumetric measurements with magnetic resonance
Mehran Azimbagirad1,2, Felipe Wilker Grillo2, Yaser Hadadian2
1University of Western Brittany, Faculty of Medicine and Health Sciences, Brest, France.
Abstract:
Purpose: Brain image volumetric measurements (BVM) methods have been used to quantify brain tissue volumes using magnetic resonance imaging (MRI) when investigating abnormalities. Although BVM methods are widely used, they need to be evaluated to quantify their reliability. Currently, the gold-standard reference to evaluate a BVM is usually manual labeling measurement. Manual volume labeling is a time-consuming and expensive task, but the confidence level ascribed to this method is not absolute. We describe and evaluate a biomimetic brain phantom as an alternative for the manual validation of BVM. Methods: We printed a three-dimensional (3D) brain mold using an MRI of a three-year-old boy diagnosed with Sturge-Weber syndrome. Then we prepared three different mixtures of styrene-ethylene/butylene-styrene gel and paraffin to mimic white matter (WM), gray matter (GM), and cerebrospinal fluid (CSF). The mold was filled by these three mixtures with known volumes. We scanned the brain phantom using two MRI scanners, 1.5 and 3.0 Tesla. Our suggestion is a new challenging model to evaluate the BVM which includes the measured volumes of the phantom compartments and its MRI. We investigated the performance of an automatic BVM, i.e., the expectation-maximization (EM) method, to estimate its accuracy in BVM. Results: The automatic BVM results using the EM method showed a relative error (regarding the phantom volume) of 0.08, 0.03, and 0.13 ( uncertainty) percentages of the GM, CSF, and WM volume, respectively, which was in good agreement with the results reported using manual segmentation. Conclusions: The phantom can be a potential quantifier for a wide range of segmentation methods.
More Related Videos
07:57Viscoelastic Characterization of Soft Tissue-Mimicking Gelatin Phantoms using Indentation and Magnetic Resonance Elastography
Published on: May 10, 2022
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Magnetic Resonance Imaging
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...