An evaluation of temperature stability and resistance in neonatal ventilator circuits
Jennifer A Ruppert1, Neil L McNinch2, Teresa A Volsko3
1Department of Respiratory Care, Akron Children's Hospital, Akron, OH, USA.
Background:
Gas conditioning minimizes complications associated with invasive ventilation of neonates. Poorly conditioned gas contributes to humidity deficit, facilitates condensate pools, and contributes to safety events. The specific aim was to objectively quantify the temperature drop across the unheated portion of a neonatal circuit and the impact condensation has to resistance to flow in the ventilator circuit.
Methods:
Ventilator circuits and filters were obtained, assembled according to manufacturer recommendations, and operational verification procedures were performed prior to data collection. A neonatal test lung was connected to each Servo-I ventilator with the following settings: pressure control IMV mode; inspiratory pressure: 14 cm H2O to achieve an exhaled tidal volume of 6.0 mL; PEEP: 5 cm H2O; pressure support: 5 cm H2O, FIO2: 0.21; set frequency 40/min; and inspiratory time: 0.4 s. The Fisher and Paykel MR850 and ChonchaTherm Neptune heaters were set at a temperature of 40°C. To evaluate both systems under similar conditions, the ChonchaTherm Neptune heater humidity control was set to midline. Heaters were turned on simultaneously and given 1 h to equilibrate. Readings for room temperature, airway temperature at the patient connection, airway resistance, exhaled tidal volume, and direct observation of circuit condensation and (or) pooling were recorded hourly for a 48-h period. Summary statistics were calculated for the variables of interest.
Results:
Mean (±SD) air temperature was 26.3°C (±1.4) for the Fisher & Paykel MR850 system and 26.2°C (±1.5), for the ChonchaTherm Neptune system. Mean (±SD) airway resistance was 229.3 cm H2O/L/s (±81.0) for the Fisher & Paykel system and 196.2 cm H2O/L/s (±39.4) for the ChonchaTherm Neptune system. Mean (±SD) tidal volume for the Fisher & Paykel MR850 system was 6.5 mL (±0.4), and for the ChonchaTherm Neptune system was 7.2 mL (±0.6).
Conclusion:
Circuit condensate increased tidal volume delivery and airway resistance. Temperature at the patient connection was lower than the temperature monitored by the system 12 inches distally, which can negatively impact gas conditioning.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...