A Validated, Automated, 3-Dimensional Method to Reliably Measure Tibial Torsion
Joanna M Stephen1,2, Robert A Teitge3, Andy Williams1,2
1Department of Mechanical Engineering, Imperial College London, London, UK.
The American Journal of Sports Medicine
|February 3, 2021
Summary
This study introduces an automated method for measuring tibial torsion, offering superior reliability and accuracy compared to traditional techniques. This advancement aims to standardize assessments and improve clinical management of lower limb conditions.
Area of Science:
- Orthopedics
- Medical Imaging
- Biomechanical Engineering
Background:
- Tibial torsion, a twist in the tibia, is crucial for lower limb function.
- Abnormal tibial torsion is linked to various painful clinical syndromes.
- Existing measurement methods lack standardization, with reported normal values varying significantly.
Purpose of the Study:
- To develop and validate an automated computer method for calculating tibial torsion.
- To compare the reliability of the automated method against current clinical techniques.
- To assess differences in torsion measurements between imaging modalities and sexes.
Main Methods:
- A controlled laboratory study involving 56 cadaveric and patient tibiae (CT scans) and 12 tibiae (MRI and structured light scans).
- Three blinded assessors applied two current clinical methods and the newly defined automated method.
- Statistical analysis included intraclass correlation coefficients (ICC) and t-tests to compare measurements.
Main Results:
- The automated method demonstrated excellent reliability (ICC = 1).
- Current clinical methods showed good reliability (ICC = 0.78-0.81).
- No significant differences were found between torsion values obtained from CT, MRI, and structured light scans.
Conclusions:
- Automated software-based measurements provide consistent, efficient, valid, and accurate results.
- Manual measurements are limited by assessor variability and experience.
- The study recommends adopting this automated method for clinical practice to establish reference values.