Objectives:
Intestinal electrical stimulation (IES) has been proposed for treating diabetes; however, its parameters need to be further systematically optimized. This study aimed to optimize the parameters of IES and investigate its possible mechanisms involving glucagon-like peptide-1 (GLP-1) in diabetic rats.
Materials And Methods:
Thirty-six high-fat diet-induced diabetic rats were chronically implanted with a pair of bipolar electrodes at the duodenum for IES. The oral glucose tolerance test (OGTT) was performed in a number of sessions with IES using different parameters and biphasic charge-balanced waveforms to derive the best values for train on-time, pulse frequency, and pulse width. Incretin hormones such as GLP-1 were assessed and the GLP-1 antagonist Exendin 9-39 was used to assess the role of GLP-1 in the ameliorating effect of IES on hyperglycemia.
Results:
The most effective IES parameters in reducing blood glucose (BG) during the OGTT were derived: 1.2 sec on, 0.3 sec off, 80 Hz, 3 msec. IES with these parameters reduced BG level by at least 29% from 15 min to 180 min (p < 0.05 for all points, N = 10). IES with these stimulation parameters increased plasma GLP-1 level at 30 min, 60 min, 90 min and gastric inhibitory peptide (GIP) level at 30 min (N = 8). Exendin 9-39 blocked the inhibitory effect of IES on BG (p > 0.05, IES + Exendin 9-39 vs sham-IES, N = 8).
Conclusion:
IES with the most effective parameters derived in this study improves hyperglycemia in diabetic rats. The ameliorating effect of IES on hyperglycemia is attributed to the enhanced release of GLP-1. IES has great potential for treating diabetes.
Related Concept Videos
Glucagon-like Receptor Agonists
GLP-1, when administered in high doses intravenously, triggers insulin secretion, inhibits glucagon release, slows gastric emptying, reduces food intake, and restores normal insulin secretion. However, its rapid inactivation by...
Hypoglycemia and Glucagon
Glucose Homeostasis: Pancreatic Islets and Insulin Secretion
Insulin and C-peptide are...
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...
Oral Hypoglycemic Agents: Glinides
Feedback Loops