First experience of efficacy and radiation exposure in 320-detector row CT fluoroscopy-guided interventions
Shota Yamamoto1, Tomohiro Matsumoto1, Satoshi Suda1
1Department of Radiology, Tokai University Hachioji Hospital, Tokai University School of Medicine, 1838 Ishikawa-machi, Hachioji, Tokyo, Japan.
Objective:
We investigated the efficacy and exposure to radiation in 320-detector row computed tomography fluoroscopy-guided (CTF-guided) interventions.
Methods:
We analysed 231 320-detector row CTF-guided interventions (207 patients over 2 years and 6 months) in terms of technical success rates, clinical success rates, complications, scanner settings, overall radiation doses (dose-length product, mGy*cm), patient doses of peri-interventional CT series, and interventional CT (including CTF), as a retrospective cohort study. The relationships between patient radiation dose and interventional factors were assessed using multivariate analysis.
Results:
Overall technical success rate was 98.7% (228/231). The technical success rates of biopsies, drainages, and aspirations were 98.7% (154/156), 98.5% (66/67), and 100% (8/8), respectively. The clinical success rate of biopsies was 93.5% (146/156). All three major complications occurred in chest biopsies. The median total radiation dose was 522.4 (393.4-819.8) mGy*cm. Of the total radiation dose, 87% was applied during the pre- and post-interventional CT series. Post-interventional CT accounted for 24.4% of the total radiation dose. Only 11.4% of the dose was applied by CTF-guided intervention. Multilinear regression demonstrated that male sex, body mass index, drainage, intervention time, and helical scan as post-interventional CT were significantly associated with higher dose.
Conclusion:
The 320-detector row CTF interventions achieved a high success rate. Dose reduction in post-interventional CT provides patient dose reduction without decreasing the technical success rates.
Advances In Knowledge:
This is the first study on the relationship between various interventional outcomes and patient exposure to radiation in 320-detector row CTF-guided interventions, suggesting a new perspective on dose reduction.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...