Clinical evaluation of a laboratory-developed quantitative BK virus-PCR assay using the cobas® omni Utility Channel
Alice Fritzsche1, Laura Berneking1, Dominic Nörz1
1Department of Medical Microbiology, Virology and Hygiene, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
Background:
In immunocompromised patients, BK Virus (BKV) reactivation may cause serious disease with high morbidity. Particularly for patient management after solid organ transplantation, monitoring of viral load in different clinical specimens is crucial to ensure early diagnosis and response to reactivation. In this study, we evaluated the clinical performance of a custom designed primer /probe set for detection of BKV on the cobas® 6800, a high-throughput platform, employing the open channel of the system for integration of a lab-developed test (LDT).
Materials/Methods:
A primer/probe set was optimized for the use on a high-throughput platform. Clinical performance was assessed in EDTA-plasma, serum and urine samples. Limit-of-detection (LOD) was determined by using a dilution series of BKV WHO standard. A CE-labeled PCR test (Altona Diagnostics) was used as a comparison to the assay.
Results:
The LOD for the LDT BKV assay was 6.7 IU/mL. Inter-and intra-run variability (at 5 x LOD) was low (<1.5 Ct in all specimens). All quality control panel specimens (Instand Germany n = 19) were correctly identified. Of 290 clinical samples tested, results were concordant for 280 samples. Sensitivity and specificity of the assay were 96 % and 98 % respectively. The quantitative analysis revealed a strong correlation (linear regression) between the CE-labelled comparator assay and the new BKV LDT assay with r2 = 0.96 for n = 123 urine samples and r2 = 0.98 for n = 167 plasma/serum samples.
Conclusion:
Compared to a CE-IVD assay, the adapted LDT showed good analytical and clinical sensitivity and specificity for the detection and quantification of BKV in different clinical specimens. It represents a convenient solution to automate the LDT workflow with low hands-on time and thus facilitates high-throughput screening for BKV reactivation in immunocompromised patients.