[Application of the Virtual Reality-Pattern Visual Evoked Potential in Forensic Visual Acuity Evaluation].
W Y Tang1, J M Chen2, R R Wang3
1Basic Medical College of Jiamusi University, Jiamusi 154007, Heilongjiang Province, China.
Fa Yi Xue Za Zhi
|February 7, 2021
Summary
This study found that virtual reality-pattern visual evoked potential (VR-PVEP) testing in monocular mode offers higher amplitude and reliability for forensic visual acuity evaluation compared to binocular mode. This method improves efficiency and overcomes fixation issues.
Area of Science:
- Ophthalmology
- Neuroscience
- Forensic Science
Background:
- Visual acuity evaluation is crucial in forensic clinical settings.
- Traditional methods may be limited by factors like poor fixation.
- Pattern Visual Evoked Potential (PVEP) offers an objective measure.
Purpose of the Study:
- To evaluate a virtual reality-based PVEP (VR-PVEP) method for forensic visual acuity assessment.
- To compare the efficacy of monocular versus binocular stimulation modes in VR-PVEP.
- To determine the suitability of VR-PVEP for reliable and efficient clinical application.
Main Methods:
- Thirty-four volunteers underwent induced refractive myopia.
- Visual acuity was set to normal in one eye and moderately impaired in the other.
- VR-PVEP tests were conducted in both monocular and binocular modes across four spatial frequencies (8x8, 16x16, 24x24, 32x32).
Main Results:
- Statistically significant differences in P100 wave amplitude between monocular and binocular modes were observed at spatial frequencies of 16x16, 24x24, and 32x32 (P<0.05).
- The amplitude of the P100 wave was consistently higher in monocular mode compared to binocular mode for the same eye.
- A moderate correlation was found between P100 wave amplitudes in monocular and binocular modes.
Conclusions:
- VR-PVEP testing, particularly in monocular mode, enhances the reliability of visual acuity evaluations in forensic contexts.
- This method effectively addresses challenges associated with poor fixation.
- VR-PVEP significantly reduces testing time and boosts overall work efficiency in forensic identification practices.